Send inquiry
Feature:
12’’1 high resolution color TFT LCD display .
One level menu , easy to use .
Maximum 720 hours tabular and graphic trends of all parameters.
Built –in thermal recorder and two batteries capacity.
Suitable for adult , pediatric, neonatal patient .
13 type arrhythmic analysis , ST segment analysis , pacemaker detection.
Anti-ESU , anti-defibrillator.
oxyCRG dynamic view and short trend coecist display , nurse call system .
standard network interface , can connect to the central monitor system .
Standard:
12.1 LCD , 5 ECG leads , RESP , NIBP,PR , SPO2, 2-TEMP , Rechargeable battery .
Optional :
Thermal recorder , 2-IBP, ETCO2 multi-gas , touch screen.
Specifications :
ECG
Lead mode : 3/5 leads
Gain : ×0.25 , ×0.5, ×1, ×2, auto
Heat rate : 15-350 bpm
Accuracy :±1%
Resolution : 1bpm
ST segment range : -2.0mV-+2.0mV
ARR detecting :yes
Pacemaker : yes
Electrode offset potential : ±300mV
Bandwidth
Surgery: 1Hz-15Hz
Monitor : 0.5Hz-35Hz
Diagnostic :0.05Hz-100Hz
SPO2:
Measuring and alarm range : 0%-100%
Resolution :1%
Accuracy :70%-100%±2% , 0%-69% unspecified
Actualization interval : about 1 sec.
Alarm delay : 10sec.
PULSE RATE:
Measuring and alarm range:20bpm-300bpm
Resolution :1bpm
Accutacy:2bpm
IBP:
Label :ART,PA,CVP,RAP,LAP,ICP,P1,P2
Measuring and alarm range
ART:0mmHg-300mmHg
PA:-6mmHg-120mmHg
CVP/RAP/LAP/ICP:-10mmHg-40mmHg
P1/P2:-10mmHg-300mmHg
Press sensor
Sensitivity :5Uv/v/mmHg
Resolution:1mmHg
Accuracy :2% or 1mmHg ,which great
Actualization interval : about 1 secretary
RESPIRATION:
Method :impedance between R-F(RA-LL)
Base line impedance range :0kΩ-2.5 kΩ
Bandwidth :0.3 Hz-2.5Hz
Measuring range : 0rpm -150rpm
Resolution :1rpm
Accuracy :2rpm
Apnea alarm :10s -40s
NIBP:
Word mode : manual , auto ,continuous
Measuring period in STAT mode: 5Min
Pulse rate range :40bpm -240bpm
Measurement types :SYS,DIA,MEAN
Measuring range
Adult mode
SYS:40mmHg-270mmHg
DIA:10mmHg-215mmHg
MEAN:20mmHg-165mmHg
Pediatric mode
SYS:40mmHg -200mmHg
DIA:10mmHg-150mmHg
MEAN: 20mmHg-165mmHg
Neonatal mode
SYS :40mmHg-135mmHg
DIA:10mmHg-100mmHg
MEAN :20mmHg-110mmHg
Resolution pressure:1mmHg
Accuracy pressure maximum mean error :5mmHg
Maximum standard deviation :8mmHg
Overpressure protection
Adult mode :2973mmHg
Prediatric mode :2403mmHg
Neonatal mode :1473mmHg
TEMPERATURE:
Channel:2
Measuring and alarm range :0℃-5℃
Resolution :0.1 ℃
Accuracy : 0.1 ℃
Actualization interval :about 1 sec
Average time constant:<10sec